My watch list
my.chemeurope.com  
Login  

Field desorption



  Field desorption (FD)/field ionization (FI) refers to an ion source for mass spectrometry first reported by Beckey in 1969.[1] In field ionization, a high-potential electric field is applied to an emitter with a sharp surface, such as a razor blade, or more commonly, a filament from which tiny "whiskers" have formed. This results in a very high electric field which can result in ionization of gaseous molecules of the analyte. Mass spectra produced by FI have little or no fragmentation. They are dominated by molecular radical cations M+. and less often, protonated molecules [M+H]^+\,.

In FD, the analyte is applied as a thin film directly to the emitter, or small crystals of solid materials are placed onto the emitter. Slow heating of the emitter then begins, by passing a high current through the emitter, which is maintained at a high potential (e.g. 5 kilovolts). As heating of the emitter continues, low-vapor-pressure materials get desorbed and ionized by alkali metal cation attachment.

Many earlier applications of FD/FI to analysis of polar and nonvolatile analytes such as polymers and biological molecules have largely been supplanted by newer ionization techniques. However, FD/FI remains one of the only ionization techniques that can produce simple mass spectra with molecular information from hydrocarbons and other particular analytes. The most commonly encountered application of FD/FI at the present time is the analysis of complex mixtures of hydrocarbons such as that found in petroleum fractions.

The recently developed liquid injection FD ionization (LIFDI) [2] technique "presents a major breakthrough for FD-MS of reactive analytes" [3]: Transition metal complexes are neutral and due to their reactivity, do not undergo protonation or ion attachment. They benefit from both: the soft FD ionization and the safe and simple LIFDI transfer of air/moisture sensitive analyte solution. This transfer occurs from the Schlenk flask to the FD emitter in the ion source through a fused silica capillary without breaking the vacuum.

References

  1. ^ Beckey H.D. Field ionization mass spectrometry. Research/Development, 1969, 20(11), 26
  2. ^ HB Linden, Liquid injection field desorption ionization: a new tool for soft ionization of samples including air sensitive catalysts and non-polar hydrocarbons, Eur. J. Mass Spectrom. 2004, 10, 459-468
  3. ^ JH Gross, Mass Spectrometry - A Textbook, Springer 2004, page 362

Bibliography

  • Alexandrov, M. L.; Gall, L. N.; Krasnov, N. V.; Nikolaev, V. I.; Pavlenko, V. A.; Shkurov, V. A., Dokl. Akad. Nauk SSSR 1984, 277, 379-383. (in Russian)
  • Prókai, László (1990). Field desorption mass spectrometry. New York: M. Dekker. ISBN 0-8247-8303-4. 
 
This article is licensed under the GNU Free Documentation License. It uses material from the Wikipedia article "Field_desorption". A list of authors is available in Wikipedia.
Your browser is not current. Microsoft Internet Explorer 6.0 does not support some functions on Chemie.DE